Breast CT: Does it Replace Mammogram or MRI?

Breast CT (breast computed tomography) is a relatively new, compression-free, three-dimensional breast imaging technique. However, breast CT is not yet a clinically validated alternative to mammography or breast MRI for routine breast cancer screening. Its availability for diagnostic breast imaging should not be confused with regulatory approval or clinical endorsement for screening.

Unfortunately, recent mainstream media coverage may have given some women the impression that breast CT can replace their regular mammogram. This has prompted many questions from my patients about its accuracy, safety and suitability for breast cancer screening.

In this article, we explore the benefits and limitations of dedicated breast CT, how it compares with mammography and breast MRI, and its potential role in breast imaging.

Table of Contents

    What is dedicated breast CT?

    Dedicated breast computed tomography (breast CT or BCT) is an imaging technology that uses X-rays to produce detailed, three-dimensional images of the breast.

    There are currently two notable dedicated breast CT technologies available internationally, including in Australia:

    • Koning Breast CT: Uses cone-beam CT technology to produce three-dimensional breast images.

    • AB-CT nu:view: Uses spiral CT technology combined with a photon-counting image detector.

    Woman lying down prone on a dedicated breast CT scanner.

    Unlike conventional mammography, breast CT does not require compression of the breast. This is particularly appealing for women who experience significant pain or discomfort during mammography.

    During the examination, a woman lies face down on a specially designed table, allowing each breast to hang naturally through an opening. The scanner rotates around the breast, capturing images from multiple angles to create a three-dimensional reconstruction. This is different from a conventional chest CT scan and is specifically designed for breast imaging.

    What are the limitations of dedicated breast CT?

    Although compression-free breast imaging may be appealing to many women, it is important to understand the limitations of dedicated breast CT before considering it as a replacement for routine screening mammography. Patient comfort is important, but it should not come at the expense of reliable and accurate breast cancer detection.

    1. Incomplete visualisation of posterior breast tissue

    Illustration of breast prone position during CT Breast. Posterior lesions may be missed as the sit outside the imaged area.

    One of the most important limitations of dedicated breast CT is its ability to adequately visualise breast tissue close to the chest wall. Due to the scanner design and patient positioning, some posterior breast tissue may fall outside the scanning field, potentially resulting in missed abnormalities.[1]

    In contrast, both mammography and breast MRI generally provide better coverage of posterior breast tissue.

    Comparing MMG and CT Breast for detection of calcifications. Two patients with posterior calcifications were missed by CT Breast.

    (A) A 51-year-old woman with a cluster of 4 microcalcifications on MMG where a group of 4 microcalcification particles were visible in the middle to posterior third of the breast. Only 1 of the calcifications was seen in the CT scan images. (B) A 65-year-old woman with a cluster of microcalcifications depicted on MMG but not seen on the CT scan images. [1]

    2. Limited assessment of the axilla (armpit)

    Assessment of the axillary lymph nodes is an important component of breast cancer evaluation, as these lymph nodes are often the first site of breast cancer spread.

    Dedicated breast CT does not adequately visualise the axilla, and abnormalities in the axillary tail (the breast tissue extending towards the armpit) may also fall outside the scanned area. Therefore, additional targeted ultrasound will be required for a comprehensive assessment.

    In contrast, both mammography and breast MRI provide some information about the axillary breast tissue and lymph nodes, with targeted ultrasound used to further investigate any suspicious findings.

    3. Suboptimal characterisation of microcalcifications

    Microcalcifications are tiny calcium deposits within the breast that can be an early sign of breast cancer, particularly ductal carcinoma in situ (DCIS). The shape and appearance of these calcifications are important in distinguishing benign from potentially cancerous abnormalities.

    In a 2026 prospective study involving 126 women with breast cancer, photon-counting CT failed to detect 44% of suspicious microcalcifications identified on conventional mammography.[2] This highlights the continued importance of mammography in detecting early signs of breast cancer, even with advances in CT imaging technology.

    4. Need for iodinated contrast material

    To achieve optimal breast cancer detection, particularly in women with dense breasts, dedicated breast CT generally requires the intravenous administration of iodinated contrast material. Without contrast, its ability to detect breast cancer has been shown to be substantially inferior to that of mammography and contrast-enhanced breast MRI.[3]

    The need for iodinated contrast limits the suitability of breast CT in patients with a history of contrast allergy, significant kidney dysfunction or certain thyroid conditions, particularly hyperthyroidism.

    In contrast, conventional mammography and 3D mammography (tomosynthesis) do not require intravenous contrast. Breast MRI uses a non-iodinated, gadolinium-based contrast agent and can be safely performed in many patients who cannot receive iodinated contrast, subject to individual medical assessment.

    5. Lack of dynamic contrast enhancement information

    Unlike standard contrast-enhanced breast MRI, routine contrast-enhanced breast CT does not provide detailed information about how quickly contrast enters and washes out of a breast lesion. This additional information, known as enhancement kinetics, helps radiologists distinguish between benign and potentially cancerous abnormalities. Combined with its excellent soft-tissue resolution, this is one of the key reasons why breast MRI remains the gold standard for contrast-enhanced breast imaging and the most sensitive imaging modality for breast cancer detection.

    6. Radiation Exposure

    Unlike breast MRI, dedicated breast CT uses ionising radiation. Although breast CT is often promoted as a low-dose imaging technique, it is important to consider the radiation exposure of a complete contrast-enhanced examination, rather than a single CT acquisition.

    The following table compares approximate radiation doses from different breast imaging modalities:

    Comparison of Radiation Doses in Breast Imaging
    Breast imaging modality Mean glandular dose per breast (mGy)
    2D mammography ~4
    3D mammography (tomosynthesis) ~5
    Contrast-enhanced mammography (CEM) ~6
    CEM combined with 3D mammography ~11
    Contrast-enhanced cone-beam breast CT (Koning) 8–26
    Contrast-enhanced photon-counting breast CT (nu:view) 9–19*
    Breast MRI 0

    *Estimated dose assuming two acquisitions (pre- and post-contrast).

    Radiation doses are indicative and may vary depending on breast size, density, equipment and imaging protocols.

    Another important consideration is that women diagnosed with breast cancer following breast CT will generally still require mammography as part of their preoperative assessment and surgical planning, particularly to evaluate microcalcifications and determine the extent of disease. Therefore, rather than replacing mammography, breast CT may become an additional imaging investigation, substantially increasing the patient's cumulative radiation exposure.

    This is an important consideration when selecting an imaging modality for breast cancer screening, particularly in younger women who require regular imaging over many years.

    7. Lack of validation for breast cancer screening

    Dedicated breast CT has not yet been adequately validated as a replacement for mammography or breast MRI in routine breast cancer screening.

    In Australia, both dedicated breast CT systems (Koning and AB-CT nu:view) are registered with the Therapeutic Goods Administration (TGA) for diagnostic breast imaging in women with breast symptoms, suspicious clinical findings or abnormal screening examinations. Their TGA-listed intended purposes do not include routine breast cancer screening in asymptomatic women.

    Large-scale clinical studies are required to establish their suitability for breast cancer screening, including cancer detection rates, missed cancers, false-positive findings, unnecessary biopsies and long-term screening outcomes.

    Until such evidence becomes available, dedicated breast CT should not replace established, evidence-based breast cancer screening modalities.

    8. Limited local clinical experience

    Dedicated breast CT is a relatively new imaging technology in Australia, with limited local clinical experience compared with established modalities such as mammography, ultrasound and breast MRI.

    Accurate breast cancer diagnosis and treatment rely not only on the technology itself but also on the training, experience and expertise of both the radiologists interpreting the images and the breast surgeons using them to guide surgical planning. Like any new imaging modality, dedicated breast CT has a learning curve.

    Therefore, the promising results reported in published studies from experienced international research centres may not necessarily translate into equivalent diagnostic performance in routine Australian clinical practice. Further local training, clinical experience and validation are required.

    9. Cost and Medicare rebates

    There is currently no dedicated Medicare item number for breast CT in Australia. Before undergoing breast CT, patients should have a clear understanding of the potential out-of-pocket costs, including those associated with any additional investigations that may be required, such as mammography, ultrasound and image-guided breast biopsy.

    Our recommendation: Should you have a breast CT scan?

    While dedicated breast CT is an interesting and evolving technology with the advantage of compression-free imaging, its current limitations and lack of robust clinical evidence prevent us from recommending it as a replacement for established breast imaging modalities.

    At this stage, we do not recommend dedicated breast CT for routine breast cancer screening in asymptomatic women or as a first-line imaging investigation in women presenting with breast symptoms. However, it may have a role as a complementary diagnostic tool in selected clinical circumstances as determined by your breast surgeon.

    Breast cancer screening should not be a one-size-fits-all approach. The most appropriate imaging modality, age to commence screening and frequency of screening should be determined by each woman's individual breast cancer risk, breast density, age and personal circumstances.

    The Personalised Breast Cancer Screening Program at Melbourne Comprehensive Breast Care (MCBC) provides an online platform where women can calculate their individual breast cancer risk and receive a free personalised screening recommendations, including the most appropriate imaging modality and screening frequency, in just 5–10 minutes.

    Patient comfort is important, but it should never come at the expense of accurate and reliable breast cancer detection.

    Frequently Asked Questions About Breast CT

    References

    1. Classification of Microcalcifications: An Intraindividual Comparison With Digital Breast Tomosynthesis. Invest Radiol, 2024. 59(12): p. 838-844.

    2. Fu, Q., et al., Performance of Photon-counting CT for Assessing Pretreatment Breast Cancer: Comparison with Mammography, MRI, and (18)F-FDG PET/CT. Radiology, 2026. 320(3): p. e253531.

    3. Wetzl, M., et al., Detectability of Breast Cancer in Dedicated Breast CT Compared With Mammography Dependent on Breast Density. Invest Radiol, 2024. 59(12): p. 861-865.

    Next
    Next

    Breast Cancer During Pregnancy: Diagnosis, Surgery and Treatment